Vulnerability Database

313,825

Total vulnerabilities in the database

CVE-2025-14096

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure:

Attacker requires physical access to the analyzer.

Temporary work Around: Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:

Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

  • Published: Dec 17, 2025
  • Updated: Dec 18, 2025
  • CVE: CVE-2025-14096
  • Severity: High
  • Exploit:

CVSS v3:

  • Severity: High
  • Score: 8.4
  • AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H